Ligand Exchange Columns for Saccharides: SUGAR S, USPpak

Saccharides are separated by combining different separation modes, such as ligand exchange, size exclusion, ion exclusion and partition/adsorption. Suitable for separation of sugar and sugar alcohol.

(Standard Columns)
Product name Counter
ion
Separation
mode
Exclusion limit
(Pullulan)
Plate number
(TPN/column)
Size(mm)
ID x Length
SUGAR SH1011
SUGAR SH1821
SUGAR SH-G
H+
H+
H+
SEC+IEX
SEC+IEX
Guard column
1,000
10,000
-
> 17,000
> 17,000
-
8.0 x 300
8.0 x 300
6.0 x 50
SUGAR SC1011
SUGAR SC1821
SUGAR SC-LG
Ca2+
Ca2+
Ca2+
SEC+LEX
SEC+LEX
Guard column
1,000
10,000
-
> 13,000
> 13,000
-
8.0 x 300
8.0 x 300
6.0 x 50
SUGAR SP0810
SUGAR SP-G
Pb2+
Pb2+
SEC+LEX
Guard column
1,000
-
> 11,000
-
8.0 x 300
6.0 x 50
SUGAR SC1211
SUGAR SC-G
Ca2+
Ca2+
N&R+LEX
Guard column
-
-
> 5,500
-
6.0 x 250
4.6 x 10
SUGAR SZ5532
SUGAR SZ-G
Zn2+
Zn2+
N&R+LEX
Guard column
-
-
> 5,500
-
6.0 x 150
4.6 x 10
USPpak MN-431 Ca2+
SEC+LEX
-
> 4,000
4.0 x 250
Product
name
Particle size
(micron)
Max. usable
pressure
(MPa)
Recommended
flow rate
(mL/min)
Max.
flow rate
(mL/min)
Usable
temp.
(deg-C)
In-column
solvent
SH1011
SH1821
SH-G
6
6
10
5
5
-
0.5 to 1.0
0.5 to 1.0
0.5 to 1.0
1.5
1.5
-
< 95
< 95
< 95
H2O
H2O
H2O
SC1011
SC1821
SC-LG
6
6
10
4
4
-
0.5 to 1.0
0.5 to 1.0
0.5 to 1.0
1.5
1.5
-
< 95
< 95
< 95
H2O
H2O
H2O
SP0810
SP-G
7
10
3
-
0.5 to 1.0
-
1.0
-
< 95
< 95
H2O
H2O
SC1211
SC-G
6
10
4
-
0.5 to 1.0
-
1.0
-
< 95
< 95
25%CH3CN/H2O
25%CH3CN/H2O
SZ5532
SZ-G
6
6
4
-
0.5 to 1.0
-
2.0
-
< 80
< 80
70%CH3CN/H2O
70%CH3CN/H2O
MN-431 - 3.5 - 0.6(1) 30 to 85
H2O
*0.6mL/min: when column temperature range is 51 to 85 deg-C.
0.5mL/min: when column temperature range is 31 to 50 deg-C.
0.3mL/min: when column temperature range is below 30 deg-C.
Separation mode:
SEC=Size exclusion, IEX=Ion exchange, LEX=Ligand exchange, N&R=Normal & reversed phase
Exclusion limit:
Measuered using pullulan standards.
Combination of one type of column with another:
When separation cannot be satisfactorily performed, use in series of an additional column of the same or a different type will improve separation. Such combination, however, is possible only among SP0810 and columns of the SC and KS series, and they must always be arranged in the following order.
SUGAR KS column + SUGAR SC column + SUGAR SP0810
When a guard column is used for any combination of columns, it must be one designed for use with the first column in the combination.
Pretreatment of sample:
The sample is usually dissolved in the eluent. Pass the dissolved sample through a 0.45micron membrane filter and inject it into the column. Use of Shodex DT disposable filter unit is recommended. Any proteins contained in the sample must be removed before filtering.
Samples containing heavy metal ions, surfactants (emulsifiers) or oily substances that adsorb on the packing should be passed through a mass of H-type cation exchange resin before filtering.
Organic acids can be removed by passing the sample through a mass of anion-exchange resin. In the case of columns of the SH group, however, samples containing organic acids require no pretreatment. Alkaline samples must be neutralized before injection; otherwise, column performance will deteriorate.
Eluent conditions:
The water used as the eluent must be deionized water that has been passed through a 0.45micron membrane filter.
Before use, it must also be degassed without fail to prevent bubbling in the column and at its outlet, because all columns of this series are heated being used. Use of degassing devices of the Shodex DEGAS is recommended.

SH type:
Although deionized water can be used alone as the eluent, adsorption of alkalis on the packing material will change the chromatogram. Addition of sulfuric or phosphoric acid to the mobile phase is, therefore1 recommended to prevent that change.
SC type and SP0810:
Basically, deionized water is used as the eluent. A polar organic solvent, if added to the eluent, will improve separation of sugar alcohols and reduce hydrophobic adsorption. The concentration of the polar organic solvent must, however, be 20% maximum.
SC type:
Addition of a salt to the eluent will improve separation of ionic substaces. The salt must be either Ca(NO3)2 or CaSO4, and its pH, in the 3 to 7 range. Also, in the case of analysis of samples containing large quantities of heavy metals, addition of about 10 to 50 ppm of Ca-EDTA will prevent deterioration of column performance.
SC1211 and SZ5532:
The eluent used with SC1211 and SZ5532 is basically a mixture of water and an organic solvent, generally acetonitrile, and it is used in partition chromatography.