Shodex: Packed Columns for HPLC Analysis |
|
| |
|
|
|
|
|
|
|
|
|
| In-column solvent | Exclusion limit (Polystyrene) |
Plate number (TPN/column) |
Particle size (micron) |
Size(mm) ID x Length |
|---|---|---|---|---|
| Toluene | ||||
| GPC HT-803 GPC HT-804 GPC HT-805 GPC HT-806 GPC HT-806M GPC HT-807 GPC HT-G |
70,000 400,000 4,000,000 (20,000,000) (20,000,000) (200,000,000) Guard column |
> 7,000 > 7,000 > 7,000 > 7,000 > 7,000 > 4,000 - |
13 13 13 13 13 18 13 |
8.0 x 300 8.0 x 300 8.0 x 300 8.0 x 300 8.0 x 300 8.0 x 300 8.0 x 50 |
| GPC UT-802.5 GPC UT-806M GPC UT-807 GPC UT-G |
20,000 (20,000,000) (200,000,000) Guard column |
> 4,400 > 4,400 > 3,300 - |
30 30 30 30 |
8.0 x 300 8.0 x 300 8.0 x 300 8.0 x 50 |
| GPC AT-806MS GPC AT-G |
(20,000,000) Guard column |
> 6,000 - |
12 15 |
8.0 x 250 8.0 x 50 |
| In-column solvent | Usable temp. range (deg-C) |
Flow rate (mL/min) | Max. usable pressure (MPa) |
|
|---|---|---|---|---|
| Toluene | Recommended | Max. | ||
| HT-803 HT-804 HT-805 HT-806 HT-806M HT-807 HT-G |
100 to 140 100 to 140 100 to 140 100 to 140 100 to 140 100 to 140 - |
0.5 to 1.0 0.5 to 1.0 0.5 to 1.0 0.5 to 1.0 0.5 to 1.0 0.5 to 1.0 - |
1.5 1.5 1.5 1.5 1.5 1.5 - |
1.5 1.5 1.5 1.5 1.5 1.5 1.5 |
| UT-802.5 UT-806M UT-807 UT-G |
100 to 210 100 to 210 100 to 210 - |
0.3 to 1.0 0.3 to 1.0 0.3 to 1.0 - |
1.5 1.5 1.5 - |
1.0 1.0 1.0 - |
| AT-806MS AT-G |
RT to 140 - |
0.5 to 1.0 - |
1.5 - |
20 - |