Ion Chromatography Columns

There are two types of ion chromatography: non-suppressor method and suppressor method. While the suppressor method requires a suppressing module, the non-suppressor method can be used for ion analysis by simply connecting an electro-conductivity detector to an all-purpose LC system. At Shodex, we provide columns for cation/anion analysis for both non-suppressor and suppressor methods. We also provide columns for analyzing transition metal ions and rare earth metal ions using the post-column reaction system. Column selection can thus be made based on your analysis needs and type of ions.

IC NI-424
IC I-524A
These columns are for anion analysis using the non-suppressor method. NI-424 is a high performance version of I-524A and can analyze fluoride ion and phosphoric acid simultaneously which was not possible with I-524A.
IC Y-521
IC YK-421
These columns are for cation analysis using the non-suppressor method. Y-521 is suitable for the separation of various alkylamines. YK-421 can be used for simultaneous analysis of monovalent and divalent cations.
IC SI-90 4E
IC SI-50 4E
IC SI-52 4E
New IC SI-91 4C
These columns are for anion analysis using the suppressor method. They demonstrate good performance in fluoride ion assays, separation/removal of foreign organic acids and are not affected by carbonate system peaks. They are thus suitable for anion analysis of various samples. They can be used with ion chromatography systems equipped with a suppressor.
IC T-521
IC R-621
T-521 is for analyzing transition metal ions and R-621 is for analyzing rare earth metal ions. The high sensitivity of metal ions in the post column reaction system can be used for measurement.

(Standard Columns)
[for Non-suppressor System]
Product name Plate number
(TPN/column)
Size(mm)
ID x Length
Applications
IC NI-424
IC NI-G
> 5,000
Guard column
4.6 x 100
4.6 x 10
Anions (high performance version of I-524A)
IC I-524A
IC IA-G
> 2,000
Guard column
4.6 x 100
4.6 x 10
Anions
IC Y-521
IC Y-G
> 3,000
Guard column
4.6 x 150
4.6 x 10
Cations (for general purpose)
IC YK-421
IC YK-G
> 2,800
Guard column
4.6 x 125
4.6 x 10
Simultaneous separation of monovalent and divalent cations

[for Suppressor System]
Product name Plate number
(TPN/column)
Size(mm)
ID x Length
Applications
IC SI-90 4E
IC SI-90G
> 5,000
Guard column
4.0 x 250
4.6 x 10
Anions (for general purpose ) (packed in PEEK column)
IC SI-50 4E
IC SI-90G
> 10,000
Guard column
4.0 x 250
4.6 x 10
Anions (for special purpose) (packed in PEEK column)
IC SI-52 4E
> 14,000
4.0 x 250 Oxyhalides ions (packed in PEEK column)
IC SI-91 4C
IC SI-90G
> 2,500
Guard column
4.0 x 100
4.6 x 10
For analysis of halogen oxides by post-column method (using PEEK column)

[For Post-column Reaction System]
Product name Plate number
(TPN/column)
Size(mm)
ID x Length
Applications
IC T-521
IC T-G
> 3,000
Guard column
4.6 x 150
4.6 x 10
Transition metal ions (packed in PEEK column)
IC R-621
IC R-G
> 1,000
Guard column
6.0 x 50
4.6 x 10
Rare earth metal ions

Product
name
Particle size
(micron)
Max. usable
pressure
(MPa)
Max. usable
flow rate
(mL/min)
Usable
temp. range
(deg-C)
In-column solvent
NI-424
NI-G
-
-
6
-
1.2
-
20 to 60
20 to 60
8mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris
+ 2mM Phenylboronic acid + 0.005mM CyDTA
I-524A
IA-G
12
12
2.5
-
2.0
-
RT to 50
RT to 50
2.5mM Phthalic acid
Y-521
Y-G
12
12
5
-
3.0
-
RT to 70
RT to 70
4mM Nitric acid
YK-421
YK-G
5
5
15
-
1.5
-
RT to 60
RT to 60
5mM Tartaric acisd + 1mM Dipicolinic acid
+ 1.5g/L Boric acid
SI-90 4E
SI-90G
9
9
10
-
2.0
-
20 to 60
20 to 60
1.8mM Na2CO3 + 1.7mM NaHCO3
SI-50 4E
SI-90G
5
9
15
-
0.8
-
20 to 60
20 to 60
3.2mM Na2CO3 + 1.0mM NaHCO3
SI-52 4E 5
5
15
1.0
18 to 40

SI-91 4C
SI-90G
9
9
5
-
2.0
-
20 to 60
20 to 60
1.8mM Na2CO3 + 1.7mM NaHCO3
T-521
T-G
12
12
-
-
-
-
-
-
3mM Nitric acid
R-621
R-G
5
5
-
-
1.5
-
-
-


(Line Filter)
IC FL-1
IC FL-1 filter
Line Filter for ion chromatography(general purpose)
This cannot be used as replacement Y-521L.
Holder and 1pc of filter
Filter (2.5mmID sintered etal, 5pcs)
IC FL-2
IC FL-2 filter
Non-metal line filter Holder and 1pc of filter
Filter (4 pcs)
IC Y-521L Exclusively for monovalent cation analysis
with Y-521.
8mmID x 50mm Length
For ion chromatography, the use of a line filter in addition to a column is highly recommended.

Products | Precautions for use | Technical info | Troubleshooting
Shodex Home | Thomson Home | Site Map | Contact us

© 2005 Thomson Instrument Co. and Shodex, Inc.