| (Features) * Small internal volume: solvent replacement is easy and quick. High performance: degassing performance is unsurpassed despite the small internal volume. * Unique safety features: internal pressure and leak monitors work to prevent damage in unusual situations. * Use with any HPLC pump; can even be used with quick-return type pumps due to low internal flow resistance. * High quality: ERC degassers are designed, manufactured and tested in an ISO 9001 certified facility. All units also conform to CE safety specifications (EMC and LVD). * Non-metallic solvent contact pans. |
| (Principle) As the eluent flows through the degassing tube made of special macromolecular film, the dissolved gas will be removed through the tube wall when the ambient pressure is reduced. This pressure reduction is controlled by the built-in pressure sensor, controller and membrane vacuum pump. (1) Membrane (2) Vacuum chamber (3) Pressure sensor (4) Controller (5) Three-way valve (6) Vacuum pump (7) Eluent (8) HPLC pump |
(Specifications)
| |
ERC-3115alpha | ERC-3215alpha | ERC-3315alpha | ERC-3415alpha |
|---|---|---|---|---|
| Solvent channels |
1 channel
|
2 channels
|
3 channels
|
4 channels
|
| Max. flow rate |
20mL/min/channel (25deg-C deionized water)
|
|||
| Degassing capacity |
Air bubble should not be observed at the outlet. (Conditions) Eluent: Deionized water Flow rate: 3mL/min Temperature: 25deg-C (Reference) Oxygen concentration at the outlet was 2ppm under the above conditions. |
|||
| Internal volume |
7mL/channel
|
|||
| Solvent contactparts |
PEEK (Inlet/Outlet port), PTFE (Membrane and tubing)
|
|||
| Displays Functions |
Power ON/OFF | "POWER" LED turns on when power is on. | ||
| Status monitor | "READY" LED turns on when the pressure is lower than the designated value. |
|||
| Pressure monitor | "PRES." LED turns on when the pressure value is higher than the designated value for the designated time. |
|||
| Leakage monitor | "LEAK" LED turns on when the leakage in the device
is detected. |
|||
| Self cleaning | Cleaning of vacuum tubing by incoming air. | |||
| Mode selection | NORM: For ordinary use. CONT: For high performance degassing. |
|||
| Outputs |
Open connector signal is output when "PRES." or "LEAK"
LED turns on.
|
|||
| Dimensions |
W71 x D310 x H136mm
|
|||
| Weight |
ca 5.0kg
|
ca 5.1kg
|
ca 5.4kg
|
ca 5.5kg
|
| Power source |
AC 100 to 240 V, 50/60Hz (Free power source)
|
|||
| Power consumption |
10VA
|
|||