Durability against Alkaline Solvents

Since the packing material of Asahipak NH2P-50 column is polymer, it is durable against alkaline solvents. To the contrary, silica-based amino columns are not durable against alkaline solvents.

Testing procedure
(1)Equilibrium(1hr) Conditions before testing
 Eluent       : 100mM Ammonium acetate(pH9.3)
 Flow rate    : 0.1mL/min
 Column temp. : 25deg-C
(2)Testing(160hrs)
 Eluent       : 5mM NaOH(pH11.4)/CH3CN=25/75
 Flow rate    : 0.1mL/min
 Column temp. : 30deg-C
(3)Replacement(0.5hr)
 Eluent       : H2O
 Flow rate    : 0.5mL/min
 Column temp. : 25deg-C
(4)Equilibrium(1hr)
 Eluent       : 100mM Ammonium acetate(pH9.3)
 Flow rate    : 0.5mL/min
 Column temp. : 25deg-C


Comparison between before and after testing
SampleBefore testingAfter testing
Elution volume
(mL)
Plate number Elution volume
(mL)
Plate number
1.Fructose7.527100 7.628100
2.Glucose9.636100 9.687700
3.Sucrose13.418400 13.409700
4.Maltose16.167200 16.087400


Column       : Shodex Asahipak NH2P-50 4E (4.6mmID*250mm)
Eluent       : CH3CN/H2O=75/25
Flow rate    : 1.0mL/min
Detector     : Shodex RI
Column temp. : 30deg-C

Products | Precautions for use | Technical info | Troubleshooting
Shodex Home | Thomson Home | Site Map | Contact us