Polystyrene Standards (21) (THF, Chloroform and DMF)

Multisolvent GPC column, Asahipak GF-HQ series can be used with various kinds of eluent. The comparison of polystyrene separation with THF, chloroform and DMF eluent is shown below. (Be careful that frequent replacement of in-column solvent may affect column life.)



Sample :
1. PS, Polystyrene (MW: 1,090,000)
2. PS, Polystyrene (MW: 190,000)
3. PS, Polystyrene (MW: 37,900)
4. PS, Polystyrene (MW: 2,980)
5. PS, Polystyrene (MW: 500)
Columns      : Shodex Asahipak GF-310 HQ + GF-510 HQ (7.6mmID*300mm each)
Eluent       : (A): THF, (B): Chloroform, (C): DMF
Flow rate    : 0.5mL/min
Detector     : Shodex UV(254nm)
Column temp. : 30deg-C