Effect of pH on Retention Time (NI-424)

When using IC NI-424 and the pH of the eluent is increased on the alkaline side, overall elution shortens. At a pH of 4.5 or higher, separation of F- and H2PO4- is difficult.


Sample : 30micro-L
1. H2PO4-
2. F-
3. Cl-
4. NO3-
5. SO42-
6. I-
7. SCN-

Column       : Shodex IC NI-424 (4.6mmID*100mm)
Eluent       : *18mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *2CyDTA(pH4.2)
Flow rate    : 1.0mL/min
Detector     : Shodex CD
Column temp. : 40deg-C
*1 The prescribed pH can be obtained by mixing the reagents in the composition described here.
*2 CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid