Carbonate and Other Anions (2) (NI-424)
IC
NI-424 has high theoretical plate number (more than 5,000). Carbonate and other anions can be separated well with NI-424.

Sample : 50micro-L
1. 40.0mg/L
HCO3-
2. 4.0mg/L
H2PO4-
3. 0.3mg/L
F-
4. 0.5mg/L
Cl-
5. 1.2mg/L
NO2-
6. 2.0mg/L
Br-
7. 2.0mg/L
NO3-
8. 2.0mg/L
SO42-
Column : Shodex IC NI-424 (4.6mmID*100mm)
Eluent : *18mM 4-Hydroxybenzoic acid + 2.8mM Bis-Tris + 2mM Phenylboronic acid + 0.005mM *2CyDTA(pH4.2)
Flow rate : 1.0mL/min
Detector : Shodex CD
Column temp. : 40deg-C
*1: The prescribed pH can be obtained by mixing the
reagents in the composition described here.
*2: CyDTA: trans-1,2-Diaminocyclohexane-N,N,N',N'-tetraacetic acid