Shodex: Packed Columns for HPLC Analysis

shodex home Home Products Technical information Support Contact us
search page shodex home page

Comparison with ODS Column (1)

The comparison of Asahipak GF-310 HQ ( a multisolvent GPC column ) and an ODS column when they were used for the measurement of POE distribution of octylphenol ethoxylate is shown here. When GF-310 HQ was used, good chromatograms were obtained using 35 and 40% acetonitrile concentration. On the other hand, when an ODS column was used, POE distribution can be measured using 55% acetonitrile concentration but the separation is not sufficient.


Sample :
Octylphenol ethoxylate

Column       : Shodex Asahipak GF-310 HQ (7.6mmID*300mm)
Eluent       : CH3CN/H2O
Flow rate    : 0.6mL/min
Detector     : Shodex UV(280nm)
Column temp. : 30deg-C
Column : ODS column from the other manufacturer Eluent : CH3CN/H2O Flow rate : 0.8mL/min Detector : Shodex UV(280nm) Column temp. : 30deg-C

Products | Precautions for use | Technical info | Troubleshooting
Shodex Home | Thomson Home | Site Map | Contact us

© 2005 Thomson Instrument Co. and Shodex, Inc.