Shodex: Packed Columns for HPLC Analysis

shodex home Home Products Technical information Support Contact us
shodex home page search page

Total Cyanide in River Water

Total cyanide in river water was analysed using DEAE-825 (a column for weak anion exchange chromatography) and an electrochemical detector (ECD).


Sample : 100micro-L of distilled river water (water)

Column       : Shodex IEC DEAE-825 (8.0mmID*75mm)
Eluent       : (10mM Sodium carbonate + 1mM Ethylenediamine)/CH3OH=90/10
Flow rate    : 1.0mL/min
Detector     : ECD(Electrode; Ag, 0mV SCE)
Column temp. : 40deg-C

Products | Precautions for use | Technical info | Troubleshooting
Shodex Home | Thomson Home | Site Map | Contact us