Shodex: Packed Columns for HPLC Analysis

shodex home Home Products Technical information Support Contact us
shodex home page search page

Cyanate and Carbamyl Phosphate

Cyanate and carbamyl phosphate were analyzed using Asahipak ES-502N 7C (a column for weak cation exchange chromatography).


Sample : 20micro-M each, 50micro-L
1. Carbamyl phosphate
2. Cyanate

Column       : Shodex Asahipak ES-502N 7C (7.6mmID*100mm)
Eluent       : 0.15M NaCl aq.
Reagent      : (I); 0.1M acetate buffer(pH3.0) containintg 2%(w/v) Phenol 
                    and 10mg/dL Sodium nitropreusside 
               (II); 1.0%(w/v) Sodium hydroxide containing 0.1%(w/v) Sodium hypochlorite
                (for hydrolysis under acidic condition/indophenol reaction)
Flow rate    : 1.0mL/min
Detector     : VIS(635nm)
Column temp. : 20deg-C
I.Koshiishi et al., Anal.Sci., 2 (1986) 81.

Products | Precautions for use | Technical info | Troubleshooting
Shodex Home | Thomson Home | Site Map | Contact us